| Switch |
Name and Switch Setting |
| 01 |
Beeper Tone |
| |
00 - off |
| |
01 - soft |
| |
02 - medium |
| |
03 - loud* |
| 04 |
RS232 Handshaking |
| |
00 - off* |
| |
01 - cts |
| |
02 - busy |
| |
03 - xon |
| 05 |
Auto transmit |
| |
00 - off* |
| |
01 - on |
| 06 |
Decimal point |
| |
00 - off |
| |
01 - on* |
| 07 |
Response data separator modes |
| |
00- space |
| |
01 - comma* |
| |
02 - tab |
| |
03 - carriage return |
| |
04 - carriage return & line feed |
| |
05 - line feed |
| 08 |
Response delimiter in RCI mode (see 12CF) |
| |
00 - carriage return |
| |
01 - carriage return & line feed* |
| |
02 - line feed |
| 09 |
Echo settings |
| |
00 - off* |
| |
01 - on |
| 0A |
Decimal Precision |
| |
00 - . |
| |
01 - .0 |
| |
02 - .00* |
| |
03 - .000 |
| |
04 - .0000 |
| |
05 - .00000 |
| 0F |
Show data labels |
| |
00 - off* |
| |
01 - on* |
| 12 |
RS-232 Protocol modes |
| |
00 - RCI* |
| |
01 - ICP |
| 13 |
Measurement modes |
| |
00 - Static measurement mode |
| |
01 - Dynamic measurement mode* |
| 14 |
Number of samples to average per static reading |
| |
0x01* |
| |
. |
| |
. |
| |
0xff |
| 15 |
Number of readings taken in static measurement |
| |
0x01* |
| |
. |
| |
. |
| |
0xff |
| 16 |
Illuminant/Observer types |
| |
00 - A_2 |
| |
01 - A_10 |
| |
02 - C_2 |
| |
03 - C10 |
| |
04 - D50_2* |
| |
05 - D50_10 |
| |
06 - D65_2 |
| |
07 - D65_10 |
| |
08 - F2_2 |
| |
09 - F2_10 |
| |
0A - F7_2 |
| |
0B - F7_10 |
| |
0C - F11_2 |
| |
0D - F11_10 |
| |
0E - F12_2 |
| |
0F - F12_10 |
| |
------------- |
| |
10 - D75_2 (added after firmware v8212) |
| |
11 - D75_10 (added after firmware v8212) |
| |
12 - D55_2 (added after firmware v8212) |
| |
13- D55_10 (added after firmware v8212) |
| 17 |
Density status types |
| |
00 - Status_A |
| |
01 - Status_E |
| |
02 - Status_I |
| |
03 - Status_T* |
| |
04 - Spectral |
| |
------------ |
| |
05 - Spectral_X (added after firmware v8212) |
| |
06 - Status_HIFI (VCMYRGBO) (added after firmware v8212) |
| |
08 - Status_Tx (added after firmware v8212) |
| |
09 - Status_Ax (reflectance) (added after firmware v8212) |
| |
0A - Status_Ex (added after firmware v8212) |
| |
0B - Status_G (added after firmware v8212) |
| |
0C - Status_M (added after firmware v8212) |
| |
0D - Status_Axt (transmission) (added after firmware v8212) |
| |
0E - Status_Mx (added after firmware v8212) |
| Note: Status responses with the sub-x option are responses with correction factors entered to more closely match the existing population of filtered densitometers. Status responses without the sub-x option correspond directly to the ANSI definitions. |
| 18 |
Data Output types |
| |
00 - Density |
| |
01 - Plus Dot |
| |
02 - Minus Dot |
| |
03 - Reflectance* |
| |
04 - Lab |
| |
05 - XYZ |
| |
06 - Yxy |
| |
07 - xyY |
| |
08 - LCh |
| |
09 - Density (VCMYRGBO) |
| 19 |
Instrument mode (added after firmware v8212) |
| |
00 - Reflectance* |
| |
01 - Transmission |
| |
|
| |
Additional Commands |
| CM |
Calibrate Drive motor |
| CR |
Calibrate Reflectance |
| |
This executes reflection calibration using the supplied calibration reference |
| CU |
Calibrate Unit (added after firmware v8212) |
| |
This enters the CR, CM and, if transmission mode is present CT |
| DM |
Drive Motor - turns on/off the drive motor |
| |
02 DM returns current motor speed |
| |
1.20DM will set the motor to the default motor speed |
| RI |
Reset Instrument to its power - up state |
| RM |
Request Measurement Initiates the measurement process |
| SV |
Send Version returns the model and firmware version |
| XE |
Clear and transmit errors |