-DensiEye Calibration Schedule

Calibration (paper white base)

The calibration is to be checked once a month and the unit recalibrated if necessary. The density deviations between reference value and the measured value must be within ± 0.01.

Calibration (absolute white base)

With this operating mode, the calibration is to be checked once a week, or whenever the ambient temperature has changed by more than 10°C / 18°F, and the unit recalibrated if necessary.

Note: The density standard and the filter type, detailed on the calibration card, must correspond with the device type specifications.